UJI BANDING PENGUAT PULSA SISTEM SPEKTROMETER GAMMA TENNELEC TC 244

Nugroho Luhur, NL and Subiharto, Subiharto and Unggul Hartoyo, UH (2015) UJI BANDING PENGUAT PULSA SISTEM SPEKTROMETER GAMMA TENNELEC TC 244. PROSIDING SEMINAR NASIONAL TEKNOLOGI DAN APLIKASI REAKTOR NUKLIR. ISSN 978-979-8500-67-1

[thumbnail of 22-nugroho_fix 140-146.pdf]
Preview
Text
22-nugroho_fix 140-146.pdf

Download (567kB) | Preview

Abstract

UJI BANDING PENGUAT PULSA SISTEM SPEKTROMETER GAMMA TENNELEC TC 244. Telah dilakukan pengujian tiga buah modul penguat pulsa sistem spektrometer gamma (amplifier spectrometer) buatan TENNELEC jenis sama yaitu TC 244. Pada perangkat spektrometer gamma penguat pulsa berfungsi untuk memperbesar amplitudo tegangan keluaran detektor radiasi (pengaturan coarse gain dan fine gain) dan berfungsi untuk membentuk atau mengolah pulsa menjadi spektrum energi menjadi semi gaussian (pengaturan shaping time dan pole-zero concelation) agar penguat pulsa dapat bekerja secara optimum. Sebelum penguat pulsa pada perangkat spektrometer gamma dipergunakan untuk melakukan analisis, penguat pulsa harus di setting/set-up pada unjuk kerja yang optimum, karena salah satu parameter yang mempengaruhi resolusi deteksi adalah pengaturan shaping time. Pengaturan shaping time yang cepat atau lambat berpengaruh terhadap lebar dan sempitnya spektrum yang dihasilkan oleh penguat pulsa yang akan berdampak pada tinggi pulsa menjadi bertambah atau berkurang yang akhirnya akan berpengaruh pada resolusi deteksi atau daya pisah (FWHM = Full Width at Half Maximum). Uji banding dilakukan dengan cara melakukan pengukuran sumber standar Co-60 pada energi 1332 keV dengan memvariasi shaping time pada penguat pulsa dan mengamati resolusi energi pada spektrum yang dihasilkan, sedangkan penguatan pada coarse gain dan fine gain diatur sehingga keluaran penguat pulsa menghasilkan amplitudo sebesar 10 Volt. Diperoleh skala set-up shaping time untuk penguat pulsa S/N 2724 pada posisi 1,5 μs dengan FWHM 2,0 keV, untuk penguat pulsa S/N 2403 pada posisi 4 μs dengan FWHM 2,0 keV dan untuk penguat pulsa S/N 3303 pada posisi 4 μs dengan FWHM 1,9 keV. Dari tiga buah modul penguat pulsa buatan TENNELEC type TC 244 diperoleh setting (set-up) shaping time yang berbeda untuk menghasilkan kinerja dan resolusi yang optimum walaupun dari produk dan jenis penguat pulsa yang sama.
Kata kunci: penguat pulsa, amplifier, shaping time
ABSTRACT COMPARATIVE TESTING AMPLIFIER SPECTROMETER TENNELEC TC 244 OF GAMMA SPECTROMETER SYSTEM. Comparative testing three amplifier spectrometer TENNELEC same types TC 244 modules of gamma spectrometer system have been done. In the spectrometer gamma system, amplifier serves to increase the output voltage amplitude of the radiation detector (setup coarse and fine gain gain) and serves to establish or cultivate pulses into the energy spectrum into a semi-gaussian (setup shaping time and pole-zero concelation) so that the amplifier can work at its optimum spectrometer. Before amplifiers spectrometer on gamma spectrometer device used to perform the analysis, amplifier spectrometer should be setup at optimum performance, because one of the parameters that affect the resolution of detection is shaping regulation time. Setup shaping time that sooner or later affect the broad and narrow spectrum produced by the amplifier spectrometer that will impact the pulse height be increased or decreased which ultimately will affect the resolution of the detection or separation power Full Width at Half Maximum (FWHM). Comparative tests conducted by measuring the standard sources Co-60 at an energy 1332 keV by varying the shaping time the amplifier spectrometer and observe the energy resolution in the spectrum produced, while strengthening the coarse gain and fine gain is set so that the output amplifier spectrometer generates an amplitude of 10 Volts , Retrieved scale setup time for the amplifier spectrometer for S/N 2724 on a position of 1.5 μs at 2.0 keV FWHM, for amplifiers spectrometer S/N 2403 on position 4 μs with a FWHM of 2.0 keV and a spectrometer amplifier S/N 3303 at position 4 μs with a FWHM of 1.9 keV. Of the three artificial amplifier spectrometer TENNELEC TC 244 type modules obtained setup shaping a different time to produce optimum performance and resolution although of products and the same type of amplifier spectrometer.
Keywords: pulse amplifier, amplifier, shaping time

Item Type: Article
Subjects: Taksonomi BATAN > Reaktor Nuklir > Teknologi Reaktor > Sistem Reaktor
Taksonomi BATAN > Reaktor Nuklir > Teknologi Reaktor > Sistem Reaktor
Divisions: BATAN > Pusat Reaktor Serba Guna
IPTEK > BATAN > Pusat Reaktor Serba Guna
Depositing User: Administrator Repository
Date Deposited: 29 Oct 2018 04:00
Last Modified: 31 May 2022 04:52
URI: https://karya.brin.go.id/id/eprint/4097

Actions (login required)

View Item
View Item