Gatot, Wurdiyanto and Hermawan, Candra and Agungagus, Budiman (2019) THE ABSOLUTELY STANDARDIZATION METHODS OF 153Sm FOR CALIBRATING NUCLEAR MEDICINE INSTRUMENTS IN INDONESIA. Journal of Applied Physical Science International, 10 (5). pp. 241-246. ISSN 2395-5260
WURDIYANTO105MsJAPSI4365.pdf
Download (466kB) | Preview
Abstract
The Standardization absolutely of radioactive sources 153Sm to calibrate the nuclear medicine equipment had been carried out in PTKMR-BATAN. This is necessary because the radioactive sources used in the field of nuclear medicine has a very short half-life in other that to obtain a quality measurement results require special treatment. Besides that, the use of nuclear medicine techniques in Indonesia develop rapidly. All the radioactive sources were prepared by gravimetric methods using KERN ABT 220-5DM type balance, traceable to SI. Standardization of 153Sm has been carried out 4phi beta (LS) – gamma coincidence methods, while the impurities has been carried out by gamma spectrometry methods which calibrated using 133Ba, 152Eu, and 241Am. The result of absolutely measurement was 88.48 kBq/g with 1.36 % of the expanded uncertainty, in covered
factor, k = 2. This value to be used for calibrating a Capintec CRC-7BT radionuclide calibrator which is a
secondary standard instrument in PTKMR - BATAN. The results show that calibration factor for Capintec CRC-7BT dose calibrator is 1.05 by about 4.6% of the expanded uncertainties.
Item Type: | Article |
---|---|
Subjects: | Taksonomi BATAN > Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja > Metrologi Radiasi Taksonomi BATAN > Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja > Metrologi Radiasi |
Divisions: | BATAN > Pusat Teknologi Keselamatan dan Metrologi Radiasi IPTEK > BATAN > Pusat Teknologi Keselamatan dan Metrologi Radiasi |
Depositing User: | Users 62 not found. |
Date Deposited: | 13 Nov 2019 03:51 |
Last Modified: | 30 May 2022 08:39 |
URI: | https://karya.brin.go.id/id/eprint/9447 |