Jati Susilo, JS and Wahyu Setia Budi, WSB and Kusminarto, KSM and Yunita Intan, YI (2009) KAJIAN SPINE PHANTOM DENGAN TEKNIK RADIOGRAFI DEXA. KAJIAN SPINE PHANTOM DENGAN TEKNIK RADIOGRAFI DEXA. pp. 385-390. ISSN 0854-2910
44_Susilo dkk_385-390_OK.pdf
Download (191kB) | Preview
Abstract
ABSTRAK
KAJIAN SPINE PHANTOM DENGAN TEKNIK RADIOGRAFI DEXA. Telah dilakukan pengukuran
terhadap spine phantom buatan sendiri dengan bahan aluminum yang ada dipasaran untuk uji densitas areal
bahan Al (areal density) dengan memanfaatkan sistem radiografi DEXA (dual energy X-ray absorptiometry)
berbasis dual energi. Hasil eksperimen diperoleh data radiograf obyek spine phantom menunjukkan bahwa
sistem ini cukup dioperasikan pada tegangan anode-katode 76 kV, arus filamen 1,5 mAs, waktu eksposure
1:27 msec dan dosis 42 μG untuk menghasilkan data kuantitatif BMD. Hasil pengukruan menunjukkan
hubungan antara BMD areal dan tebal step spine phantom bersifat linear dengan linearitas Y = 0,1292X
+0,0432 dengan R2 =0,9988, sesuai dengan spine phantom standart dengan linearitas 0,0955X + 0,4275
dengan R2 = 0,9996. Hasil ini menunjukkan bahwa sistem ini diharapkan dapat dikembangkan untuk
diagnose mineral tulang dengan pesawat radiografi yang lain sehingga lebih efektif dan efisien.
Kata-kata kunci: Spine Phantom, DEXA, sinar-x, BMD
ABSTRACT
STUDY SPINE PHANTOM WITH DEXA RADIOGRAPHY TECHNIQUE. It have been done by the
measurement to the home made spine phantom with the aluminum substance on the market for the test
density areal materials Al (areal density) by exploiting the radiography DEXA system (dual energy X-ray
absorptiometry) base on dual energy. From the experiment result obtained spine phantom radiograph object
data indicate that this system is operated enough at anode-katode voltage 76 kV, filament current of 1,5 mAs,
exposure time 1:27 msec and dose of 42 μG to yield the quantitative data BMD. This home made spine
phantom can give enough the information concerning the ROI (region of interest) density areal obyek which
measured. To each presentation, obtained by some data in the form of ROI and BMD (bone mineral density)
then graph, so that obtained by the equation of graph BMD vs thick ROI which in the form of equation of
linear, similar to equation of linear graph at spine phantom plane wafting DEXA. This result indicate that
this system is expected can be developed for the diagnosis of bone mineral with plane of other radiography
so that more efficient and effective.
Key words: Spine Phantom, DEXA, X-ray, BMD
Item Type: | Article |
---|---|
Subjects: | Taksonomi BATAN > Keselamatan dan Keamanan Nuklir Taksonomi BATAN > Keselamatan dan Keamanan Nuklir |
Divisions: | BATAN > Pusat Teknologi dan Keselamatan Reaktor Nuklir IPTEK > BATAN > Pusat Teknologi dan Keselamatan Reaktor Nuklir |
Depositing User: | Administrator Repository |
Date Deposited: | 12 Nov 2018 03:24 |
Last Modified: | 31 May 2022 03:33 |
URI: | https://karya.brin.go.id/id/eprint/4905 |