Setiawan, Doni and Febrian, Muhamad Basit and Abidin, Abidin and Rattyananda, Badra Sanditya and Setiadi, Yanuar (2024) Fabrication of sealed source scandium-46 (46Sc) as an alternative calibration standard. Journal of Radioanalytical and Nuclear Chemistry. ISSN 0236-5731, 1588-2780
Full text not available from this repository. (Request a copy)Abstract
Cobalt-60 is usually used for calibration standard that nowadays, it hard to fabricated. This research aims to fabricate of Sealed Source 46Sc as alternative Calibration Standard. Sealed source 46Sc radioactivity was 1.17 ± 0.08 µCi, sealed with an aluminum label. Radiation exposure on the surface opened container is 19.27 ± 0.09 µSv/h, a closed is 10.78 ± 0.15 µSv/h and the distance of 1 m from an open container is 4.40 ± 0.04 µSv/h whereas closed has 2.40 ± 0.04 µSv/h. Multi-channel analyzer γ-spectrometry showed the 46Sc gamma spectrum similar with characteristics from standard 60Co source.