ANALISIS KEKUATAN LANDASAN ALUMINIUM PADA PERANGKAT BRAKITERAPI MEDIUM DOSERATE
Sari
A STRENGHT ANALYSIS OF THE ALUMINiUM ANVIL OF THE MEDIUM DOSE RATE (MDR) BRACHYTHERAPHY EQUIPMENT. A strenght analysis of the aluminium anvil of the medium dose rate (MDR) brachytheraphy equipment has ben performed. The analysis is to evaluate the strenght of anvil structure used in brachytheraphy equipmen. The anvil serves to connect the wire driver module, the source container module and the channels distributor module. The method is to analyze stresses by using mathematical model. The analysis shows that the aluminium anvil has the largest stress at 6,676 N/mm2 and it is still under the yeild stress at 199,73 N/mm2 . Evaluation results of the structure analysis can be used in designing the anvil for MDR brachytheraphy equipment.