UJI KARAKTERISTIK PENCAHAYAAN ILLUMINATOR RADIOGRAFI

Supano, Supano (2011) UJI KARAKTERISTIK PENCAHAYAAN ILLUMINATOR RADIOGRAFI. Widyanuklida, 11 (1). pp. 27-33. ISSN 1410-5357

[thumbnail of 03_UJI_KARAKTERISTIK_PENCAHAYAAN_ILLUMINATOR.pdf]
Preview
Text
03_UJI_KARAKTERISTIK_PENCAHAYAAN_ILLUMINATOR.pdf

Download (2MB) | Preview

Abstract

Uji Karakteristik Pencahayaan Illuminator Radiografi. Uji karakteristik
pencahayaan 5 buah illuminator (viewer) telah dilakukan. Pengujian
dilakukan untuk mengetahui karakteristik pencahayaan illuminator
radiografi industri. Karakteristik pencahayaan meliputi kecerahan
(luminansi), penyebaran (difusi), dan kerataan (uniformity). Uji karakteristik
pencahayaan mengacu pada standar SNI-18-6932-2002 menggunakan alat
ukur luxmeter EXTECH Model 409026. Hasil uji menunjukkan bahwa
illuminator memiliki kecerahan yang berbeda. Kecerahan illuminator
memenuhi syarat untuk mengamati radiograf dengan densitas kurang dari
3,0 untuk illuminator nomor 2, densitas kurang dari 3,5 untuk illuminator
nomor 1 dan nomor 5, densitas kurang dari 4 untuk illuminator nomor 3 dan
nomor 4. Hasil uji penyebaran cahaya menunjukkan semua illuminator
memiliki faktor penyebaran melebihi 0,7. Hasil uj i kerataan cahaya
menunjukkan bahwa illuminator nom or 4 tidak memenuhi persyaratan.
(Luminance Characteristics Testing Of Radiographic Illuminator. The
luminance characteristics of five industrial radiographic illuminators have
been tested, based on the Indonesian National Standard SNI-J8-6932-2002.
The luminance characteristics included brightness, diffusion, and uniformity
have been examined using luxmeter EXTECH Model 409026. The results
show that the light brightness of illuminators are slightly different.
Brightness of illlJ('lina tors are eligible to view the radiograph with density
less than 3.0 for number 2 , density less than 3.5 for number J and 5, and
density less than 4.0 for number 3 and 4. The light diffusion testing shows
that all illuminators have factor of diffusion more than O.7. The Light
uniformity testing shows that illuminator number 4 does not meet the
requirement.)

Item Type: Article
Uncontrolled Keywords: Radiografi, illuminator (viewer), uji karakteristik pencahayaan, SNI 18-6932-2002.
Subjects: Taksonomi BATAN > Isotop dan Radiasi > Pemanfaatan Isotop dan Radiasi > Bidang Industri
Divisions: BATAN > Pusat Pendidikan dan Pelatihan
IPTEK > BATAN > Pusat Pendidikan dan Pelatihan
Depositing User: Administrator Repository
Date Deposited: 12 Jun 2019 04:20
Last Modified: 02 Jun 2022 02:24
URI: https://karya.brin.go.id/id/eprint/308