Holnisar, Holnisar and Gatot, Wurdiyanto and Hermawan, Candra (2016) The Calibration Control of Capintec Dose Calibrator CRC-7BT Using 137Cs Method. 2nd International Conference on the Sources, Effects and Risks of Ionizing Radiation (SERIR2). pp. 125-128.
2016 HOLNISAR SERIR2.pdf
Download (161kB) | Preview
Abstract
To maintain accuracy and traceability of measurement results of activity measurement Capintec Dose Calibrator CRC-7BT the radionuclides specialy of short half life such as 99mTc, 131I, 18F, and others. Laboratory Standardization PTKMR-BATAN develop a calibration control method tapped characteristics 137Cs measurements on all cal settings measurement and produce calibration function equation Y = 20.74 X-0,93 with R2=0.998. With this equation the activity values obtained 99mTc , 131I , 18F in the calculation , is used as a reference value to obtain the calibration factor value of 99mTc, 131I and 18F activity by comparing the calculation values with the output activity of the measuring
devices. Differences in measurement results using the calibration correction factors between the calibration control method and calibration factor data of Capintec Dose Calibrator CRC-7BT in 2011 was still below 5 %, still within
tolerance. It shows the performance of Capintec Dose Calibrator CRC-7BT belong Laboratory Standardization is still good and the Calibration data Factor in 2011 can still be used. In order for the performance of Capintec Dose
Calibrator CRC-7BT always observed the calibration control method must be performed periodically.
Item Type: | Article |
---|---|
Subjects: | Taksonomi BATAN > Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja Taksonomi BATAN > Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja Taksonomi BATAN > Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja > Metrologi Radiasi Taksonomi BATAN > Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja > Metrologi Radiasi |
Divisions: | BATAN > Pusat Teknologi Keselamatan dan Metrologi Radiasi IPTEK > BATAN > Pusat Teknologi Keselamatan dan Metrologi Radiasi |
Depositing User: | Administrator Repository |
Date Deposited: | 16 May 2018 05:56 |
Last Modified: | 31 May 2022 04:44 |
URI: | https://karya.brin.go.id/id/eprint/1735 |