Dosimetric accuracy of the cone-beam CT-based treatment planning of the Vero system: a phantom study

Heru, Prasetio (2016) Dosimetric accuracy of the cone-beam CT-based treatment planning of the Vero system: a phantom study. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS,, 17 (4). pp. 106-113.

[thumbnail of 2016 HERU PRASETIO JURNAL OF APPLIED.pdf]
Preview
Text
2016 HERU PRASETIO JURNAL OF APPLIED.pdf

Download (682kB) | Preview

Abstract

We report an investigation on the accuracy of dose calculation based on the conebeam computed tomography (CBCT) images of the nonbowtie filter kV imaging system of the Vero linear accelerator. Different sets of materials and tube voltages were employed to generate the Hounsfield unit lookup tables (HLUTs) for both CBCT and fan-beam CT (FBCT) systems. The HLUTs were then implemented for the dose calculation in a treatment planning system (TPS). Dosimetric evaluation was carried out on an in-house-developed cube phantom that consists of waterequivalent slabs and inhomogeneity inserts. Two independent dosimeters positioned
in the cube phantom were used in this study for point-dose and two-dimensional (2D) dose distribution measurements. The differences of HLUTs from various materials and tube voltages in both CT systems resulted in differences in dose
calculation accuracy. We found that the higher the tube voltage used to obtain CT images, the better the point-dose calculation and the gamma passing rate of the 2D dose distribution agree to the values determined in the TPS. Moreover,the insert materials that are not tissue-equivalent led to higher dose-calculation inaccuracy. There were negligible differences in dosimetric evaluation between
the CBCT- and FBCT-based treatment planning if the HLUTs were generated using the tissue-equivalent materials. In this study, the CBCT images of the Vero system from a complex inhomogeneity phantom can be applied for the TPS dose calculation if the system is calibrated using tissue-equivalent materials scanned at high tube voltage (i.e., 120 kV).
PACS number(

Item Type: Article
Subjects: Taksonomi BATAN > Keselamatan dan Keamanan Nuklir
Taksonomi BATAN > Keselamatan dan Keamanan Nuklir
Taksonomi BATAN > Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja
Taksonomi BATAN > Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja
Divisions: BATAN > Pusat Teknologi Keselamatan dan Metrologi Radiasi
IPTEK > BATAN > Pusat Teknologi Keselamatan dan Metrologi Radiasi
Depositing User: Administrator Repository
Date Deposited: 16 May 2018 04:14
Last Modified: 31 May 2022 04:44
URI: https://karya.brin.go.id/id/eprint/1324

Actions (login required)

View Item
View Item